TMQ.CV Series Floor Type Pulsation Vacuum Steam Autoclave Sterilizer
Application:
With pressure steam as medium, it is used in sterilizing medical dressing, utensils, class A, Class B hollow lumen instruments, catheters, and waste in the operating room, clinic department, CSSD, biological laboratories and other healthcare sectors.
Main Features:
●PLC control technology, colored touch screen;
●Affusion, pulse vacuum, heating, sterilizing, exhaust, vacuum drying whole process are automatically controlled by program;
●Dynamic pulsant vacuum exhaust air: Vacuum up to-0.08MPa, can exhaust steam in medical dressings, Class A hollow lumen instruments;
●Pulsation behind vacuum drying system to ensure parcels dry .
●High-performance built-in steam generator and steam-water separator devices, ensure quality steam, without external steam source;
●With instruments, dressing, rubber and customized procedures for user's optional;
●With manually Bowie & Dick leaking and vacuum testing procedures;
●0.22μm HEPA filter can ensure the vacuum steam is amicrobic, to avoid further contamination;
●Dynamically LCD display temperature, pressure, time, sterilizing status, malfunction alarm;
●Equipped with maintenance reminding function. User-friendly maintenance equipment and replacement of wearing parts;
●Beep reminding when finishing;
●Mini printer is optional to real-time record the whole sterilizing process and malfunction informations.
Safety Protection:
●Over-temperature protection: Over the setting temperature, system automatically cut off power and display alarm;
●Door safety interlock devices: Only the doors closed in place, sterilizers to start working procedures. If there is dangerous pressure in the chamber, the door can not be opened;
●Dual overpressure protection: Automatic releasing safety valve: If pressure is over design , safety valve will open to release; Overpressure automatic protection control: If pressure is over setting, safety valve will automatic release and display alarm;
●Anti-dry alarm: below the water level, automatically cut off the heating power, and display alarm information;
● Electronic circuit safety devices: AC main circuit with a short-circuit protection, over-voltage DC control circuits with overload protection.
Main Technical Parameters:
Designed Pressure: -0.1/0.28MPa Max. Working Pressure: 0.23MPa
Pressure Display Precision: ± 0.1KPa
Sterilization Temperature Range: 105℃~136℃
Temperature Display Precision: ±0.1℃
Temperature uniformity: ㄑ±0.5℃
Chamber Material: SUS304 stainless steel
Model |
Volume |
Chamber Dimension |
Overall Dimension(LxWxH) |
Net Weight |
Voltage |
Power |
CLASS B-TMQ.CV 5090P 185L |
185L |
500x950mm |
1350x750x1742mm |
230Kg |
380V |
11Kw |
|