TMQ.C Series Floor Type Fully Automatic Autoclave Sterilizer
Application:
It is suitable to be used in sterilizing dressing, Class N Cavity devices, solid instruments, medical catheters, culture medium, and waste in the operating room, clinic department, CSSD, biological laboratories and other healthcare sectors.
Main Features:
●Built-in steam generator; microcomputer control;
●Affusion, Heating, sterilization, exhaust, drying whole processes are automatically controlled by program;
●The temperature, pressure, time, running status and error alarms can all be dynamically displayed;
●With instruments, instruments bags, dressings, rubber, liquid and other sterilization procedures for customized optional;
●With evaporator, external water;
●Self-check and LCD display function;
●Beep reminding when ending sterilization;
●Mini printer is optional which record the whole sterilization process.
Safety Protection:
●Over-temperature protection: Over setting temperature, system automatically cut off power;
●Door safety interlock devices: the door is only closed in place, power can be connected to heat steam; If there is pressure inside the chamber, the door will not be opened;
●Automatic releasing pressure safety valve: Over setting pressure, releasing valve opens to release the pressure;
●Anti-dry alarm: below the water level, automatically cut off the heating power, and display alarm;
● Electronic circuit safety devices: AC main circuit with a short-circuit protection, over-voltage DC control circuits with overload protection.
Main Technical Parameters:
Designed Pressure: 0.28MPa
Designed Temperature: 150℃
Max. Working Pressure: 0.23MPa
Temperature Selective Range: 105℃~136℃
Temperature Precision: ±0.1℃
Temperature uniformity: ㄑ±0.5℃
Chamber Material: SUS304 stainless steel
Model |
Volume |
Chamber Dimension |
Overall Dimension(LxWxH) |
Net Weight |
Voltage |
Power |
CLASS N-TMQ.C-6310 300L |
300L |
632x1000mm |
1400x890x1742mm |
350Kg |
380V |
16Kw |
|