280L Horizontal Cylindrical Pressure Steam Sterilizer With Drying Function
Use
The horizontal sterilizer is adopted the way of gravity exchange to releasing the cold air from the chamber more completely to ensure the reliable sterilization. The control system automatically adjust the steam inlet and outlet according to chamber temperature during sterilizing. This unit is ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed, culture media, etc.
Characteristics
•Sterilizing course automatic sterilization controlled, easy to operate.
•With drying function, suitable dressing drying.
•With over-temperature, over-pressure auto-protect device.
•The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa. It can't be start on if the door doesn't close well.
•The safety valve will be open automatically when the inner pressure over 0.24MPa and the steam be exhausting to the water tank.
•The power will be cut-off automatically, cut-off the water and alarm whem the machine lacking water.
•The chamber of the sterilizer is made of stainless steel.
Technical Data
Technical Data |
YF-150YDA |
YF-200YDA |
YF-280YDA |
YF-400YDA |
YF-500YDA |
Sterilizing chamber volume(mm) |
150L (Ø440×1000) |
200L (Ø515×1000) |
280L (Ø600×1000) |
400L (Ø700×1100) |
500L (Ø700×1300) |
Working pressure |
0.22MPa |
Working temperature |
134¡æ |
Adjustment of temperature |
40¡«134¡æ |
Time for sterilization |
0¡«60min |
Time for drying |
0¡«60min |
Heat average |
≤±2¡æ |
Power |
9KW / AC380V 50Hz |
12KW / AC380V 50Hz |
18KW / AC380V 50Hz |
Dimension (mm) |
1400×600×1300 |
1440×670×1650 |
1400×770×1780 |
1430×880×1830 |
1800×900×1820 |
Package dimension (mm) |
1550×750×1850 |
1560×820×1850 |
1680×920×2100 |
1600x1050x2100 |
1850x1050x2100 |
G.W./N.W. |
320 / 240 Kg |
350 /260 Kg |
465 /365 Kg |
530 /420 Kg |
580 /470 Kg |
|