SEARCH:
      
   
Tabletop Steam Autoclaves
Portable Stainless Steel Steam Autoclaves
Vertical Steam Autoclaves
Floor Type Steam Sterilizers
Horizontal Sterilizers
Manual Door Pulsant Vacuum Sterilizers
Mechanized Door Pulsant Vacuum Sterilizers
Cassette Autoclave Sterilizers
EO Gas Sterilizers
Transfusion Water Bath(Ventilated) Sterilizers
Rapid Cooling Injection Sterilizers
Ampoule Sterilizers
Chinese Powder Medicine Drying Sterilizers
Medical Drying Cabinets
Hot Air Sterilizers
Dynamic air sterliizers
Foodstuffs , Meats, Fruits, Vegetables Sterilizers
Clothes or tools sterilizers
Plasma Sterilizers
Low-Temperature Formaldehyde Steam Sterilizers
Ultraviolet Lamp Trolleys
Washer Disinfectors
Small Ultrasonic Cleaners
Large Ultrasonic Cleaners
OEM Ultrasonic Cleaners
Water Distillers
Medical Sterilization & Cleaning Accessory Equipments
Infrared inoculation sterilizers
Glass Bead Sterilizers
Decolorization, transfer, universal shakers
Mixer Series
Oscillator Series
UV Analyzers
Mini Centrifuges
Constant Temperature Incubators
Rotating Shaker Incubators
Mini Vacuum Pumps
Magnetic Stirrers

Products

     Floor Type Steam Sterilizers    100L 
WG-0.1JS(F)

Sliding Door Fluctuation Vacuum Sterilizer

 

1 Technical parameters

 

No

Technical parameters

Design index

1*

Vessel design life

Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles)

Inner liner 316L, jacket s30408 life 15 years / 30000 sterilization cycles

2*

Design pressure of inner chamber

-0.1~0.3MPa

3*

Jacket design pressure

0.3MPa

4

Design temperature

150

5

Maximum operating temperature

138

6

Maximum working pressure

0.25MPa

7

Opening pressure of inner chamber safety valve

0.28MPa

8*

Opening pressure of jacketed safety valve

0.28 MPa

9

Inner chamber hydrostatic test pressure

0.39MPa

10

Jacket test pressure

0.52  MPa

11

Temperature display accuracy

0.1

12

Pressure display accuracy

0.1KPa

13

Sterilization temperature control accuracy

±02

14

Temperature uniformity

±1

15

Amplitude and times of vacuum pulsation

Amplitude - 80 ~ 0kpa, times 1 ~ 99

16*

Number of pulsations

Amplitude - 80 ~ 80kPa, times 1

17*

Times of positive pressure fluctuation

1~3

18*

Ultimate vacuum

-97 KPa

19

Vacuum leakage rate

0.13KPa/min

20

surface temperature

The external surface temperature shall not exceed 25 of the ambient temperature

21

Dryness

The weight gain of fabrics after sterilization and drying shall not exceed 0.6%, and that of utensils, tools and instruments after drying shall not exceed 0.1%.

 

2 Specification and model of liftgate

 

 Model

Tank size

Volume

L)

Boundary dimension

Weight

Kg

Steam consumption

Kg/C

Power

Width x height

Depth

L x W x H

WG-0.1JS(F)

400*450

560

100

1000*1050*1800

700

14

380V,50HZ 2KW

 

The product information will probably be updated without notice, if that happens, please refer to the latest technical material.
    



Contact Information
Yongfeng Enterprise Co., Ltd.
Add: No.1, Lane 110, Shengfeng Road, Ningbo, 315010, China
Tel:  86-574-87255354
Fax: 86-574-87125894
E-mail: sales@yongfeng-medical.com  
Web Site: http://www.yongfeng-medical.com  

Copyright 2024, All rights reserved. Design By HiSupplier Online Inc.
Browse Keywords By Alphabet: A B C D E F G H I J K L M N O P Q I S T U V W X Y Z 0-9