No
|
Technical parameters
|
Design index
|
1*
|
Vessel design life
|
Internal tank s30408, jacket Q245R life 10 years (20000 sterilization cycles)
|
Inner liner 316L, jacket s30408 life 15 years / 30000 sterilization cycles
|
2*
|
Design pressure of inner chamber
|
-0.1~0.3MPa
|
3*
|
Jacket design pressure
|
0.3MPa
|
4
|
Design temperature
|
150℃
|
5
|
Maximum operating temperature
|
138℃
|
6
|
Maximum working pressure
|
0.25MPa
|
7
|
Opening pressure of inner chamber safety valve
|
0.28MPa
|
8*
|
Opening pressure of jacketed safety valve
|
0.28 MPa
|
9
|
Inner chamber hydrostatic test pressure
|
0.39MPa
|
10
|
Jacket test pressure
|
0.52 MPa
|
11
|
Temperature display accuracy
|
0.1℃
|
12
|
Pressure display accuracy
|
0.1KPa
|
13
|
Sterilization temperature control accuracy
|
±02℃
|
14
|
Temperature uniformity
|
±1℃
|
15
|
Amplitude and times of vacuum pulsation
|
Amplitude - 80 ~ 0kpa, times 1 ~ 99
|
16*
|
Number of pulsations
|
Amplitude - 80 ~ 80kPa, times 1
|
17*
|
Times of positive pressure fluctuation
|
1~3
|
18*
|
Ultimate vacuum
|
-97 KPa
|
19
|
Vacuum leakage rate
|
≤0.13KPa/min
|
20
|
surface temperature
|
The external surface temperature shall not exceed 25 ℃ of the ambient temperature
|
21
|
Dryness
|
The weight gain of fabrics after sterilization and drying shall not exceed 0.6%, and that of utensils, tools and instruments after drying shall not exceed 0.1%.
|