|
|
Products |
|
Ampoule Sterilizers XG1.SS Manual Door Ampoule Leak-hunting Sterilizers |
XG1.SDF-5.0 Manual Door Ampoule Leak-hunting Sterilizer (Double Door)
|
Ampoule sterilizers |
Type£ºXG1.SDF-5.0 Manual door ampoule leak-hunting sterilizer (Double Door) |
Made Department£ºDisinfection and sterilization equipment branch |
Overall Dimensions£º4362¡Á2975¡Á1990mm |
Net Weight£º5200kg |
|
| |
USAGE |
|
Product Introduction This series of sterilizers are mainly used for sterilization, leak-hunting and cleaning of ampoule, oral liquid, injection bottles and etc in pharmaceutical companies and medical institutions. Pressure steam is used as the sterilization medium with many spots for permutation of steam into air to increase temperature, which ensures a complete air exhaust and uniform temperature increase within 105¡æ¡«127¡æ.After sterilization, color water leak-hunting and vacuum leak-hunting can be performed on ampoule and oral liquid. These products are in full accordance with ISO9001 quality standard and GMP regulations in design, manufacture and inspection. |
| |
Specifications |
|
Technical Parameter Chamber Size: 4000¡Á1000¡Á1200mm Chamber Volume: 5.0 cubic meter Max Steam Consumption: 300kg/cycle Max Color Water Consumption: 4500kg/cycle Max Deionized Water Consumption: 1800kg/cycle Max Compressed Air Consumption: 6.5 cubic meter Dynamic Power(3-AC380V/50Hz/kw): 8.4 kW Control Power: 220V/500w |
| |
Characteristics |
|
Performance characteristic ¡öThe main body is a horizontal rectangular structure manufactured under strict quality and safety standard. ¡öThe safe, reliable airproof door easy to be manipulated. ¡öThe reliable, nice and compact pipeline system convenient for maintenance. ¡öThe strong, realistic and loading trolley and specialized structure trays. ¡öThe stable and reliable touch screen monitoring system, convenient for operation. ¡öTwo kinds of methods of leak-hunting for ampoule and one for oral liquid available for choice. ¡öExtractive stainless steel cover. ¡öGMP validation interface is equipped for validation anywhere and anytime. |
| |
|
|
|
Configuration ¡öThe inner shell is made of 304 stainless steel and polished. ¡öSlopy pneumatic valve from GEMU Germany. ¡öSupercharge pump from STERLING Germany. ¡öGermany nash_elmo Vaccum Pump. ¡öPLC+DCS remote control computer system from SIEMENS Germany. ¡öManual double door for all. ¡öMonorail system ¡öSterilizing trolley size:900x920x965mm ¡öquantity of sterilizing trolley:2 pcs ¡ö1-2ml tray size:295x300¡Á75mm ¡öquantity of 1-2ml tray:512 trays ¡öSize of 5ml sterilizing trays:295x300¡Á90mm ¡öQuantity of 5ml trays:360 trays ¡ö Size of10ml trays:295x300¡Á105mm ¡öQuantity of 10ml trays:324 trays ¡ö Size of 20ml trays: 295x300¡Á130mm ¡öQuantity of 20ml trays:252 trays ¡öLoading quantity of 1ml ampoules:336960 pcs ¡öLoading quantity of 2ml ampoules:302400pcs ¡öLoading quantity of 5ml ampoules:129600pcs ¡öLoading quantity of 10ml: ampoules : 87480pcs ¡öLoading quantity of 20ml ampoules:42320pcs | |
|
|
Standard on the product ¡öQ/XH 53-2000 |
| |
|
|
Installation requirements ¡öSteam: 0.3-0.5Mpa saturated ¡öColor water: 0.15-0.4Mpa ¡öDeionized water: 0.2-0.4Mpa ¡öCompressed air: 0.4-0.7Mpa dust removed and water, oil filtered |
|
|
|
|
|
|
¡öFirst sterilizing, then leak hunting, or first leak hunting, then sterilizing for the programs, leak hunting is done by means of both vacuum and posive pressure. Two times cleaning and two adding-pressure pumping. |
|
| |
|
The product information will probably be updated without notice, if that happens, please refer to the latest technical material. |
|
|
|
|